Temperature Data Logger Model: TR-55i-TC
Property

- Thermo Recorder with Thermocouple Type K, J, T, S Temperature Sensor.
- Possible Measurement Range : -199.9 to 1,700ºC.
- The Sensor is made to order, for the vary use (sold separetely).
- Water Resistance IP64 (rated for use in daily life).
- Logging Capacity 16,000 reading.

Usability

- Compatible with Data Collector TR-57DCi or Communication Port TR-50U2.
- Infrared Interface (with TR-57DCi).

Attribute

Specification
  TR-55i-TC TR-55i-Pt TR-55i-V TR-55i-mA TR-55i-P
Measurement Channels Temperature 1ch Temperature 1ch Voltage 1ch 4-20mA 1ch Pulse Count 1ch
Sensor

Thermocouple

Type: K, J, T, S

Pt100, Pt1000  (3 wire) - - -
Measurement Units °C, °F °C, °F V, mV mA P
Measurement Range -199 to 1700 °C -199 to 600 °C 0 to 22 V

0 to 20 mA

(Operationalup to 40 mA)

Input Signal:

・Non-voltage Contact Input

・Voltage Input (0to27V)

Detection Voltage

  Lo: 0.5 V or less

  Hi: 2.5V or more

Input Impedance:

Approx. 100 KΩ pull up

Chattering Filter:

  ON: 15 Hz or less

  OFF: 3.5 kHz or less

Accuracy(*1) Thermocouple Measurement

±0.3°C + 0.3 % rdg

[Type K, J, T]

±1°C +0.3%rdg[Type S]
Cold Junction Compensation
±0.3°C [10 to 40°C]
±0.5°C [-40 to 10°C / 40 to 80°C]

±0.3°C + 0.3 % rdg

[10 to 40°C]

±0.5°C + 0.3 % rdg

[-40 to 10°C / 40 to 80°C]

±0.5 mV + 0.3 % rdg

[10 to 40°C]

±1 mV + 0.5 % rdg

[-40 to 10°C / 40 to 80°C]

±0.05 mA + 0.3 % rdg

[10 to 40°C]

±0.1mA + 0.3 % rdg

[-40 to 10°C / 40 to 80°C]

Maximum Count

61,439 / Recording Interval
Note: The above temperatures [__°C] are for the operating environment of the Input Module.

Measurement

Resolution
Type K, J, T : 
0.1 °C
Type S : 
Approx. 0.2 °C
0.1 °C up to 400mV: at 0.1mV, 
up to 800mV: at 0.2mV, 
up to 999mV: at 0.4mV, 
up to 3.2V: at 1mV, 
up to 6.5V: at 2mV, 
up to 9.999V: at 4mV, 
up to 22V: at 10mV
0.01 mA  
Logging Capacity 16,000 readings
Recording Interval Select from 15 choices: 1, 2, 5, 10, 15, 20, 30 sec. or 1, 2, 5, 10, 15, 20, 30, 60 min.
Recording Mode Endless (Overwrite oldest data when capacity is full)
or One Time (Stop recording when capacity is full)
LCD Display Items Measurements, Recording Status, Battery Life Warning, etc.
Communication Optical Communication (proprietary protocol)
Interfaces Infrared(IR) Communication (IrPHY 1.2 low power)
Power(*2) Lithium Battery (LS14250) x 1
Battery Life(*3)

Approx. 14 months

(10 months with IR Communication Enabled)

Approx. 24 months

(14 months with IR Communication Enabled)

Approx. 16 months

(11 months with IR Communication Enabled)

Approx. 16 months

(11 months with IR Communication Enabled)

Approx.

16 to 24 months

(11 to 18 months with IR Communication Enabled)

Dimensions H 62 mm x W 47 mm x D 19 mm (excluding protrusions and Input Module)
Weight Approx. 55 g (including battery / excluding Input Module)
Operating Environment -40 to 80 °C
Waterproof Capacity(*4) IP64: Splash proof (rated for use in daily life) Note: Input Module is not water resistant.
Accessories Input Module
(TCM-3010)
Input Module
(PTM-3010)
Input Module
(VIM-3010)

Input Module
(AIM-3010)

Input Module
(PIC-3150)
Lithium Battery (LS14250), Strap, User's Manual (Warranty Included)
Data Collection Devices

Communication Port: TR-50U2, TR-50U

Data Collector: TR-57DCi

(*1) "rdg" stands for reading. 

(*2) The included lithium battery (LS14250) is not sold in stores. Please purchase the optional battery set (TR-11P2) for replacement. 

(*3) Battery life varies depending upon the ambient temperature in which it is used, the recording interval, the frequency of communication, and the battery performance.

All estimates are based on operations carried out with a new battery and are in no way a guarantee of actual battery life. When the infrared communication function is enabled,

battery life may be shortened further if the unit is used under the inverter type fluorescent lighting. 

(*4) This is the waterproof capacity of the data logger with the Input Module connected.